Andrea Rich | Datwyler Group: Do dry rubber electrodes survive earwax and artificial sweat, and how quickly does EIS reveal degradation?
11:21 - 13:49
Other snippets from this talk
Summary of the clip:
Do dry rubber electrodes survive earwax and artificial sweat, and how quickly does EIS reveal degradation?
Discs of different rubber compounds were immersed in artificial earwax and artificial sweat for one day and seven days, with EIS and electrical tissue impedance measured against the before-exposure baseline. The aim: to see how resistant each compound is to body fluids, and whether EIS can evaluate environmental exposure for smartwatches worn while running or earbud hearables.
Material A proved very resistant to earwax: before, after one day and after seven days, the EIS and ETI curves sit on top of each other. Material B was less promising β even after one day, and especially after seven days, both electrical tissue impedance and impedance spectra increased considerably. With two performant materials, selection can be based on earwax versus sweat resistance.
In artificial sweat, material A again performed very well, EIS and ETI essentially unchanged after one and seven days. Material B changed less drastically than in earwax, but still not ideal. Notably, its response after one day and after seven days is the same: whatever happens to the material happens quickly β good for screening, bad that the change appears at all.
In this short video, you can learn:
* Material A's EIS and ETI spectra stayed essentially unchanged in both earwax and sweat, even after seven days.
* Material B degraded far more in earwax than in sweat, so fluid-specific immersion separates materials.
* Material B's impedance shift was complete after one day and identical at seven days, so one-day immersion suffices for screening.
π **Clip Abstract** Rubber-compound discs were immersed in artificial earwax and sweat for one and seven days, then measured by EIS and ETI against baseline. Material A stayed unchanged in both fluids, while material B degraded more in earwax than sweat and already after one day.
About the speaker:
* Speaker: Andrea Rich
* Company: Datwyler Group
* Event: Electronic Textiles RESHAPED 2026 | Berlin
* Location: Estrel, Berlin
Electronic Textiles RESHAPED 2027 | Berlin
π
Save the date: 20 and 21 October 2027
π Berlin, Germany
π Info: https://www.techblick.com/electronictextilesreshaped
#ArtificialEarwax, #ArtificialSweat, #ImpedanceSpectroscopy, #ElectricalTissueImpedance, #WearableElectrodes, #Hearables
This is a highlight of the presentation:
From Impedance Spectra to Wearable Performance: Practical EIS Characterization of Skin-Contact Electrodes
More Highlights from the same talk.
05:38 - 07:05
What fixed sample area, thickness, amplitude and stabilization time make EIS repeatable for dry rubber electrodes?
What fixed sample area, thickness, amplitude and stabilization time make EIS repeatable for dry rubber electrodes?
The starting question was whether impedance spectroscopy can serve as a repeatable, reliable evaluation method for rubber electrodes specifically. Working with the University of Genoa, the team ran a parameter study on dry rubber electrodes, capturing both impedance magnitude and phase shift of the electrodeβtissue interface. The goal was to define the parameter space that yields repeatable measurements, and to make different materials easier to compare.
Two geometric controls matter first: the same exposed surface area in contact with the electrolyte solution and the same sample thickness. The electrical settings were a 15 millivolt final signal amplitude measured at open-circuit potential, a data point density of 10 points per decade, and a 15-minute stabilization time to let everything reach equilibrium before the scan begins. The electrolyte was 0.1 molar sodium chloride.
Timing is dominated by equilibrium, not by the measurement itself: the 15-minute stabilization is the longest part, while the scan takes only two minutes once started. The physical setup was a 3D-printed chamber, with the electrode on the right side compressed against a hole in the chamber by a spring-loaded pin that holds it in place before measurement.
In this short video, you can learn:
* Fixing exposed surface area to the electrolyte and sample thickness is what makes impedance spectra from different dry rubber materials comparable.
* A 15 millivolt amplitude at open-circuit potential, 10 points per decade and 0.1 molar sodium chloride defined the measurement window.
* The 15-minute stabilization dominates the time budget of the whole protocol, while the scan itself runs in two minutes.
π **Clip Abstract** A University of Genoa collaboration tested whether impedance spectroscopy can repeatably evaluate dry rubber electrodes, fixing exposed surface area and sample thickness so materials stay comparable. The protocol used 15 millivolts at open-circuit potential, 10 points per decade, 15 minutes of stabilization and 0.1 molar sodium chloride in a 3D-printed chamber with a spring-loaded pin.
About the speaker:
* Speaker: Andrea Rich
* Company: Datwyler Group
* Event: Electronic Textiles RESHAPED 2026 | Berlin
* Location: Estrel, Berlin
Electronic Textiles RESHAPED 2027 | Berlin
π
Save the date: 20 and 21 October 2027
π Berlin, Germany
π Info: https://www.techblick.com/electronictextilesreshaped
#DryRubberElectrodes, #ImpedanceSpectroscopy, #SodiumChlorideElectrolyte, #SpringLoadedPin, #SkinContactElectrodes, #ElectronicTextiles
09:03 - 10:32
Does electrical impedance spectroscopy at 10 Hz actually predict electrode-tissue impedance for dry rubber electrodes?
Does electrical impedance spectroscopy at 10 Hz actually predict electrode-tissue impedance for dry rubber electrodes?
Four rubber electrode types with markedly different chemistries were fully characterised using two measurements: electrical impedance spectroscopy and electrode-tissue impedance. The ETI dataset covered at least three subjects and at least three electrodes per subject. This paired approach allowed the EIS response to be compared directly with the skin-contact electrical behaviour across the four rubber materials rather than relying on a single measurement route.
At 10 Hz, plotted EIS values against ETI showed a good correlation for the four tested dry rubber electrodes. Lower charge-transfer resistance and higher effective capacitance were reflected in lower impedance. No universal truth for all dry electrodes is claimed, but for these four fully tested materials the correlation indicates EIS can be a faster route to meaningful data.
A key caveat is materials that strongly interact with water. Such a material acts like a sponge, absorbing salt water and making both impedance and electrical impedance spectroscopy very low. That low EIS did not correlate with a very low ETI. The method therefore requires careful consideration of the materials and systems under investigation before it is implemented.
In this short video, you can learn:
* At 10 Hz, EIS versus ETI showed a good correlation across four dry rubber electrode chemistries tested on at least three subjects with at least three electrodes.
* Lower charge-transfer resistance and higher effective capacitance were reflected in lower impedance, supporting EIS as a faster characterization route.
* Materials that strongly absorb salt water can show very low EIS without a correspondingly low ETI, so material-system interactions must be checked.
π **Clip Abstract** Four chemically different rubber electrodes were characterized by electrical impedance spectroscopy and electrode-tissue impedance across at least three subjects with at least three electrodes. At 10 Hz, EIS versus ETI showed a good correlation for the tested materials, but water-absorbing materials acted like sponges and produced very low EIS that did not correlate with ETI.
About the speaker:
* Speaker: Andrea Rich
* Company: Datwyler Group
* Event: Electronic Textiles RESHAPED 2026 | Berlin
* Location: Estrel, Berlin
Electronic Textiles RESHAPED 2027 | Berlin
π
Save the date: 20 and 21 October 2027
π Berlin, Germany
π Info: https://www.techblick.com/electronictextilesreshaped
#ElectrodeTissueImpedance, #ElectricalImpedanceSpectroscopy, #DryRubberElectrodes, #ChargeTransferResistance, #WearableElectronics, #SkinContactSensing
About this talk
β From Impedance Spectra to Wearable Performance: Practical EIS Characterization of Skin-Contact Electrodes Β· Andrea Rich Β· The Future of Electronics RESHAPED 2026 Berlin