Sascha Reinhardt | Instrument Systems: What field of view and resolution can 12- versus 151-megapixel camera systems deliver on microLED microdisplays?
02:47 - 04:12
Other snippets from this talk
Summary of the clip:
What field of view and resolution can 12- versus 151-megapixel camera systems deliver on microLED microdisplays?
The 12-megapixel camera system, such as LumiTop, can cover field of views up to 14 millimeters by 10 millimeters with 3.4 micrometers per pixel. It can also reach 1.4 millimeters by 1 millimeter at 0.34 micrometers per pixel. The field of view becomes really small with that system.
Switching to a 151-megapixel camera system provides the benefit of a still-large field of view. You can cover the complete micro-display with very high resolution. With 11 magnification, this becomes a very high-resolution system, which allows detailed study of individual pixels of the micro-display.
The systems let you measure individual properties of each pixel at the same time. That matters because the 12-megapixel field of view gets really small, while the 151-megapixel system keeps a large field of view at high resolution. The 11 magnification example shows how individual micro-display pixels can be studied in detail.
In this short video, you can learn:
* A 12-megapixel LumiTop system covers up to 14 millimeters by 10 millimeters at 3.4 micrometers per pixel or 1.4 millimeters by 1 millimeter at 0.34 micrometers per pixel.
* A 151-megapixel system maintains a large field of view while covering the complete micro-display at very high resolution.
* At 11 magnification, the 151-megapixel system enables detailed study of individual micro-display pixels.
📋 **Clip Abstract** Microdisplay metrology can measure individual pixel properties simultaneously. A 12-megapixel LumiTop camera system trades field of view for resolution, while a 151-megapixel system keeps a large field of view and at 11 magnification enables detailed study of individual micro-display pixels.
About the speaker:
* Speaker: Sascha Reinhardt
* Company: Instrument Systems
* Event: nnovation Festival: Additive, Printed, Sustainable Electronics, MicroLEDs and AR/VR Displays, Perovskites | July 2026
* Location: Online Platform
Annual access to our onsite event(s), all online events and a library featuring over 1800 presentations (recordings + PDF) across all our onsite and onsite events and technology areas. Join us and over 2000 members from around the world now.
Explore and join now --> https://www.techblick.com/
#LumiTop, #12MegapixelCamera, #151MegapixelCamera, #MicrodisplayPixelInspection, #MicroLEDMetrology, #DisplayMetrology
This is a highlight of the presentation:
Microled Microdisplay Metrology
More Highlights from the same talk.
06:25 - 08:36
Why does a fixed color-correction matrix fail on microLED, and how far does live spectral calibration cut the error?
Why does a fixed color-correction matrix fail on microLED, and how far does live spectral calibration cut the error?
In the first step, an RGB camera sensor in the LumiTop captures the display, and its RGB imaging is converted to XYZ using a color correction matrix based on spectral assumptions. In parallel, the spectrum is measured at a given point. Comparing center values where the spectrometer spot sits with that spectrometer measurement allows fine correction of the 2D camera’s XYZ values.
This approach works quite well as long as spectral variation is not so large as in LCDs. Better results normally come from what is called live calibration: the measured spectrum directly optimizes the color correction matrix. The matrix is not fixed anymore; it is adapted from measurement to measurement, improving the results.
With a standard matrix, the color difference is about four color points. When the current measured spectrum is used live—what is actually measured with the microdisplay—the color difference goes down to one color point in the mean. That quantifies the improvement from live spectral adaptation.
In this short video, you can learn:
* A fixed color correction matrix converts RGB camera data to XYZ using spectral assumptions, so it works well only while spectral variation is not large.
* Live calibration uses the measured spectrum to optimize the color correction matrix for each measurement instead of keeping it fixed.
* A standard matrix gave about four color points of color difference, while live current-spectrum calibration reduced it to one color point in the mean.
📋 **Clip Abstract** RGB camera imaging in the LumiTop is converted to XYZ through a color correction matrix, with a parallel spectrometer measurement used to finely correct center XYZ values. Live calibration instead optimizes the matrix from the measured spectrum, reducing color difference from about four color points with a standard matrix to one color point in the mean.
About the speaker:
* Speaker: Sascha Reinhardt
* Company: Instrument Systems
* Event: nnovation Festival: Additive, Printed, Sustainable Electronics, MicroLEDs and AR/VR Displays, Perovskites | July 2026
* Location: Online Platform
Annual access to our onsite event(s), all online events and a library featuring over 1800 presentations (recordings + PDF) across all our onsite and onsite events and technology areas. Join us and over 2000 members from around the world now.
Explore and join now --> https://www.techblick.com/
#ColorCorrectionMatrix, #LiveCalibration, #SpectrometerMeasurement, #ColorDifference, #MicroLEDDisplays, #DisplayMetrology
About this talk
→ Microled Microdisplay Metrology · Sascha Reinhardt · TechBlick Innovations Festival July 2026