Azar Alizadeh | GE HealthCare Technology & Innovation Center: How does patient mobility impact the performance and reliability of wearable medical devices, and what design considerations are necessary to address these challenges?
00:07:14 - 00:07:49
Other snippets from this talk
Summary of the clip:
How does patient mobility impact the performance and reliability of wearable medical devices, and what design considerations are necessary to address these challenges?
The speaker addresses the challenges associated with patient mobility and its impact on the performance of wearable medical devices. They note that devices that perform well when patients are bedridden may encounter difficulties when patients become mobile and are exposed to the conditions of daily life. This transition from high-acuity to mobility introduces new challenges that must be addressed.
The speaker uses the example of the COVID-19 pandemic to illustrate this point, noting that devices performed well when patients were monitored in bed but faced challenges when patients became more active. This highlights the need for devices to be robust and reliable in a variety of real-world conditions. The speaker identifies three key factors that communities like NextFlex are well-suited to address: diagnostics, cost and volume, and mobility.
The speaker emphasizes that these three factors—diagnostics, cost and volume, and mobility—are critical considerations for the development of wearable medical devices. They suggest that organizations like NextFlex, with their expertise in flexible electronics, are well-positioned to develop solutions that address these challenges and enable the widespread adoption of wearable medical devices in healthcare.
In this short video, you can learn:
* The impact of patient mobility on the performance of wearable medical devices.
* The challenges associated with transitioning from high-acuity to mobile monitoring.
* The three key factors that need to be addressed for successful wearable device development: diagnostics, cost/volume, and mobility.
📋 **Clip Abstract** This segment discusses the challenges of maintaining the performance and reliability of wearable medical devices as patients transition from bedridden to mobile states. It identifies diagnostics, cost/volume, and mobility as key areas where flexible electronics can contribute to improved device design and functionality.
🔗 Link in comments 👇
#WearableMedicalDevices, #DeviceReliability, #MobileHealthMonitoring, #FlexibleElectronics, #DigitalHealth, #SemiconductorTechnology
This is a highlight of the presentation:
Manufacturing of Low-Cost Wearable Physiological and Molecular Marker Monitoring Patches
More Highlights from the same talk.
11:22 - 13:34
How does GE design multi-modal vital sign monitoring patches to achieve clinical-grade fidelity in a disposable form factor?
How does GE design multi-modal vital sign monitoring patches to achieve clinical-grade fidelity in a disposable form factor?
The more complex generation integrates a three-lead ECG, dual respiration measurements, reflective PPG and temperature in one patch. An electronics unit controls all patch functions with Bluetooth or other wireless capabilities, accelerometry, temperature and onboard storage. Attention centers on electronics performance and optimized manufacturing matched to clinical counterparts, with head-to-head comparisons of signal quality and signal stability against clinical devices.
To turn these devices into single-use products, the manufacturing approach uses flexible hybrid electronics to be cost-effective and scalable. That requires a marriage of disparate manufacturing components and supply chains. The design and production are optimized so the patch's electronics can match clinical counterparts, while the flexible hybrid route supports disposable manufacturing rather than a reusable clinical device.
The miniaturized version removes the three-dimensional ECG, and its respiration modality also changes, but it aims to maintain the same type of clinical fidelity. Development is driven by the need based on operation, as noted in the talk. Both generations therefore trade functions and form factor while retaining careful electronics design and clinical-comparison validation.
In this short video, you can learn:
* The complex patch combines three-lead ECG, dual respiration, reflective PPG and temperature.
* Flexible hybrid electronics enable cost-effective scalable manufacturing for single-use patches.
* The miniaturized patch drops 3D ECG and changes respiration while targeting comparable clinical fidelity.
📋 **Clip Abstract** This clip examines two generations of disposable vital sign patches, focusing on a complex version with three-lead ECG, dual respiration, reflective PPG and temperature. It explains how flexible hybrid electronics, optimized electronics and head-to-head clinical comparisons support scalable single-use manufacturing while the miniaturized version changes modalities but targets clinical fidelity.
About the speaker:
* Speaker: Azar Alizadeh
* Company: GE Healthcare
* Event: Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
* Location: Online
Join TechBlick - for the leading global event on wearable electronics
https://www.techblick.com/
#FlexibleHybridElectronics, #ThreeLeadECG, #DualRespiration, #ReflectivePPG, #WearableSensors, #FlexibleElectronics
17:41 - 19:56
Why do 600 Hz heart-rate and 100 Hz PPG sampling cap a patch at 3.5 days, and what does cutting the rate buy?
Why do 600 Hz heart-rate and 100 Hz PPG sampling cap a patch at 3.5 days, and what does cutting the rate buy?
The patch is built from high-conductivity conductors that stay both bendable and stretchable over a very long period. The functional answer to fit is mechanical: the middle regions of the device stretch or compress the wiring so one unit adapts to different body sizes. A person with a thin body frame uses the same device as a larger wearer, removing the need for many different device sizes.
An Android app displays all of the parameters, and embedded algorithms convert the raw wave functions into clinically familiar metrics such as heart rate and respiration rate. Critically, the device supports continuous transmission at a very fast data rate for up to 3.5 days. That continuous link is what defines the usable wear window for subacute monitoring.
The sampling rates are stated plainly: heart rate at 600 Hz and photoplethysmography at 100 Hz. For subacute monitoring those rates are unnecessary, so transmission can be reduced and battery lifetime extended up to 1 week. The device tolerates showering, is comfortable to wear, and is being taken forward with AFRL colleagues for air medical transport.
In this short video, you can learn:
* Heart rate is sampled at 600 Hz and PPG at 100 Hz, sustaining continuous transmission for up to 3.5 days.
* Reducing the transmission rate for subacute monitoring extends battery lifetime up to 1 week.
* Stretch-and-compress regions in the middle of the wiring let a single device fit different body frames, and the work is now aimed with AFRL at air medical transport.
📋 **Clip Abstract** A stretchable, high-conductivity clinical patch that adjusts mechanically to different body frames and streams raw waveforms through an Android app into heart rate and respiration rate. Continuous transmission at 600 Hz heart rate and 100 Hz PPG runs 3.5 days, while lowering the rate for subacute monitoring reaches 1 week of battery life.
About the speaker:
* Speaker: Azar Alizadeh
* Company: GE Healthcare
* Event: Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
* Location: Online
Join TechBlick - for the leading global event on wearable electronics
https://www.techblick.com/
#StretchableInterconnects, #PPGSamplingRate, #ContinuousVitalSignTransmission, #AirMedicalTransport, #WearableSensors, #FlexibleElectronics
15:15 - 17:42
Why must printed SpO2 sensors and main motherboards be placed separately on flexible vital sign monitoring patches?
Why must printed SpO2 sensors and main motherboards be placed separately on flexible vital sign monitoring patches?
The electronics unit is split into two separate components in the more 3D DCG version because the PPG, which measures parameters such as blood oxygenation, must sit below the clavicle for good reflective SpO2. The motherboard is instead placed close to the side of the ribs for comfort. Both are highly miniaturized and copper-flex based; the motherboard supports ECG, respiration and temperature, while the PPG module remains separate.
For printed wires, the work was done closely with DuPont and manufacturing supplier GM Nameplate. The approach is a multilayer print structure that allows shielding of the wires. The base substrate is PET, where multiple inks have been optimized. This device can be bent at a very sharp angle, important because the devices are adjustable for different body sizes.
The choice of dielectrics or laminations allows these devices to be compatible with high-voltage applications, in case defibrillation is one of the hospital requirements. The result is described as very high-performing, with low- and high-conductivity devices. The separate miniaturized modules and shielded printed wires support adjustable devices for different body sizes.
In this short video, you can learn:
* PPG must be below the clavicle for reflective SpO2, while the motherboard sits near the side of the ribs for comfort.
* A multilayer print structure on PET shields printed wires, enabling sharp bending and body-size adjustment.
* Dielectric and lamination choices make the devices compatible with high-voltage hospital defibrillation requirements.
📋 **Clip Abstract** The electronics unit is split into two highly miniaturized copper-flex components: a PPG module below the clavicle for reflective SpO2 and a motherboard near the side of the ribs for comfort that supports ECG, respiration and temperature. Printed wires use a multilayer structure on PET for shielding, sharp bending and adjustable body sizes, with dielectric and lamination choices compatible with high-voltage defibrillation requirements.
About the speaker:
* Speaker: Azar Alizadeh
* Company: GE Healthcare
* Event: Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
* Location: Online
Join TechBlick - for the leading global event on wearable electronics
https://www.techblick.com/
#ReflectiveSpO2, #MultilayerPrintShielding, #PETSubstrate, #CopperFlexModules, #WearableSensors, #FlexibleElectronics
About this talk
→ Manufacturing of Low-Cost Wearable Physiological and Molecular Marker Monitoring Patches · Azar Alizadeh · The Future of Electronics RESHAPED 2025 USA
Other talks by Azar Alizadeh
- Disposable, Adjustable Clinical Grade Vital Sign Monitoring Systems · Wearable Sensors & Continuous Vital Signs Monitoring 2021
More highlights from Azar Alizadeh
- ▶ How does GE design multi-modal vital sign monitoring patches to achieve clinical-grade fidelity in a disposable form factor? · Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
- ▶ Why must printed SpO2 sensors and main motherboards be placed separately on flexible vital sign monitoring patches? · Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
- ▶ Why do 600 Hz heart-rate and 100 Hz PPG sampling cap a patch at 3.5 days, and what does cutting the rate buy? · Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
- ▶ What supply chain architecture moves flexible hybrid wearables from laboratory prototypes into relevant manufacturing environments? · Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021