Sandro Francesco Tedde | Siemens Healthcare: Why do some flat panel X-ray detectors trade sensitivity for resolution, while direct conversion keeps high resolution?

04:21 - 05:45

Other snippets from this talk

Summary of the clip:

Why do some flat panel X-ray detectors trade sensitivity for resolution, while direct conversion keeps high resolution?

This flat panel X-ray detector can detect signals even down to single X-ray photons, giving very high sensitivity. Its limitation is relatively low resolution. When an X-ray photon is absorbed, light spreads isotropically from that point. The generated photons travel across several pixels, so the optical spread degrades the detector’s potential resolution. This explains the sensitivity-versus-resolution trade-off described for this detector.

Direct conversion is a second technology, mainly based on semiconductors as X-ray absorbers, specifically amorphous selenium. Here the X-ray is converted directly into electrical charges, which are then read out. Because the charges remain where they are created—at least in the column where they are created—and are driven by the field lines generated by an applied bias, the detector retains really high resolution.

The contrast is structural: indirect detection converts X-rays to spread light before readout, while direct detection converts X-rays to localized charge. In the direct case, charge transport follows the bias-generated field lines, preserving spatial information. The clip links this retained high resolution to mammography, ending on that application without further detail.

In this short video, you can learn:
* Indirect detectors achieve single-photon X-ray sensitivity but lose resolution because absorbed photons generate light that spreads isotropically across several pixels.
* Direct conversion uses amorphous selenium as the X-ray absorber and converts X-rays directly into electrical charges for readout.
* In direct conversion, charges stay near their creation column and drift along bias-generated field lines, preserving high resolution.

📋 **Clip Abstract** This flat panel X-ray detector can detect signals even down to single X-ray photons, but light spreads isotropically across several pixels and degrades its potential resolution. Direct conversion with amorphous selenium converts X-rays directly into electrical charges that stay localized and follow bias-generated field lines, retaining high resolution and linking the advantage to mammography.

About the speaker:
* Speaker: Sandro Francesco Tedde
* Company: Siemens Healthcare
* Event: Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
* Location: Online

Join TechBlick - for the leading global event on wearable electronics
https://www.techblick.com/

#AmorphousSelenium, #DirectConversionReadout, #IsotropicLightSpread, #SinglePhotonSensitivity, #FlatPanelDetectors, #XRayImaging

This is a highlight of the presentation:

Perovskites: The Emergence of a New Era for High-Resolving and Ultra-Sensitive Flat Panel X-Ray Detectors

Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021

​

13-14 October 2021

Online

Organised By:

TechBlick

​

More Highlights from the same talk.

11:06 - 12:49

Can a single X-ray absorber cover both mammography's soft and radiography's hard spectra better than selenium or CsI?

Can a single X-ray absorber cover both mammography's soft and radiography's hard spectra better than selenium or CsI?

Detector performance hinges on high X-ray absorption, so the absorption coefficient is plotted against photon energy from 10 to 100 keV, the range in which mammography and radiography images are made. Two regions matter: the mammography window at 30 to 40 keV and the harder radiographic window at 70 to 90 keV. Each material must be selected for high absorption in the spectrum it will actually see.

Selenium, the red curve, absorbs well in the mammography region but badly in radiography, so you would never use it for RAD. Cesium iodide, the green curve, sits close to cadmium telluride: poor in the mammography spectrum, strong in the radiographic one at 70 to 90 keV. That is precisely why cesium iodide serves as the conversion material for radiographic applications.

Methylammonium lead iodide, MAPbI3, is the dark green curve. It shows a bit less absorption in the radiographic spectrum than cesium iodide and a bit less in the mammography spectrum than selenium. Neither deficit is disqualifying: in principle the perovskite is useful for both kinds of imaging, letting one absorber span soft mammography energies and hard radiographic energies.

In this short video, you can learn:
* Selenium absorbs strongly across the 30–40 keV mammography window but is effectively ruled out for the 70–90 keV radiographic spectrum.
* Cesium iodide, plotted close to cadmium telluride, absorbs poorly at mammography energies yet is the conversion material of choice for radiography.
* MAPbI3 gives a bit less absorption than either incumbent in its own window, but in principle covers both mammography and RAD with one absorber.

📋 **Clip Abstract** Plotting absorption coefficient from 10 to 100 keV explains why selenium owns mammography at 30–40 keV and cesium iodide owns radiography at 70–90 keV. Methylammonium lead iodide absorbs slightly less in each window, yet in principle serves both, replacing two conversion materials with one.

About the speaker:
* Speaker: Sandro Francesco Tedde
* Company: Siemens Healthcare
* Event: Electronic Textiles, Wearable Sensors & Printed/Flexible Sensors 2021
* Location: Online

Join TechBlick - for the leading global event on wearable electronics
https://www.techblick.com/

#MAPbI3Perovskite, #XRayAbsorptionCoefficient, #CesiumIodideConverter, #AmorphousSeleniumAbsorber, #FlatPanelXRayDetectors, #MedicalImagingPhysics


About this talk

→ Perovskites: The Emergence of a New Era for High-Resolving and Ultra-Sensitive Flat Panel X-Ray Detectors · Sandro Francesco Tedde · Printed & Flexible Sensors & Actuators 2021

More Snippets
CONTACT US

KGH Concepts GmbH

Mergenthalerallee 73-75, 65760, Eschborn

+49 17661704139

admin@techblick.com

TechBlick is owned and operated by KGH Concepts GmbH

Registration number HRB 121362

VAT number: DE 337022439

  • LinkedIn
  • YouTube

Sign up for our newsletter to receive updates on our latest speakers and events AND to receive analyst-written summaries of the key talks and happenings in our events.

Thanks for submitting!

© 2026 by KGH Concepts GmbH

bottom of page